3-12
Operation
Measurement considerations
Many measurements made with the Model 2000-SCAN-20 are subject to various effects that
can seriously affect low-level measurement accuracy. The following paragraphs discuss these
effects and ways to minimize them.
Path isolation
Path isolation is simply the equivalent impedance between any two test paths in a
measurement system. Ideally, the path isolation should be infinite, but the actual resistance and
distributed capacitance of cables and connectors results in less than infinite path isolation values
for these devices.
Figure 3-4
Path isolation resistance
Path isolation resistance forms a signal path that is in parallel with the equivalent resistance
of the DUT, as shown in
. For low-to-medium device resistance values, path isolation
resistance is seldom a consideration; however, it can seriously degrade measurement accuracy
when testing high-impedance devices. The voltage measured across such a device, for example,
can be substantially attenuated by the voltage divider action of the device source resistance and
path isolation resistance, as shown in
. Also, leakage currents can be generated
through these resistances by voltage sources in the system.
R
DUT
E
DUT
R
PATH
V
R
IN
DUT
Scanner
Card
DMM
= Source Resistance of DUT
= Source EMF of DUT
= Path Isolation Resistance
= Input Resistance of Multimeter
R
DUT
E
DUT
R
PATH
R
IN
Summary of Contents for 2000-SCAN-20
Page 11: ......
Page 12: ...1 General Information General Information ...
Page 17: ...1 6 General Information ...
Page 18: ...2 CardConnections andInstallation CardConnec tionsand Installation ...
Page 35: ...2 18 Card Connections and Installation ...
Page 36: ...3 Operation Operation ...
Page 51: ...3 16 Operation ...
Page 52: ...4 Service Information Service Information ...
Page 63: ...4 12 Service Information ...
Page 64: ...5 Replaceable Parts Replace ableParts ...
Page 67: ...5 4 Replaceable Parts ...
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ...A Specifications Section A Specs 7 8 03 4 32 PM Page 1 ...
Page 79: ......
Page 80: ......