Model 2000-SCAN Scanner Card User's Manual
Section 3: Operation
2000-SCAN-900-01 Rev. A / April 2018
3-7
Setting up an RTD measurement on the front panel
To set up an RTD measurement on the front panel:
1. After selecting the channels for your scan, select
Temperature
.
2. From the SETTINGS tab, select the Transducer options.
3. Select
2-Wire RTD
,
3-Wire RTD
, or
4-Wire RTD
.
4. Below the Transducer options, the selected transducer type will have RTD options. Select the
appropriate RTD type.
5. Next to Unit, use the selection button to choose between Celsius, Fahrenheit, or Kelvin.
6. Set the NPLC to
1
.
7. Select the
SCAN
tab.
8. Set the scan count.
9. Set the scan interval.
10. You can now start your scan in one of two ways:
Select
Start
on the SCAN screen.
Press the
TRIGGER
key, and select a measurement state. Select
Initiate Scan
to start the
scan.
Typical applications
Resistor testing
The 2000-SCAN can be used to test up to 10 resistors using 2-wire measurements or up to five
resistors using 4-wire measurements. These tests use the DMM6500 2-wire and 4-wire resistance
functions.