No comm link error
A “No Comm Link” error indicates that the front panel processor has ceased communication
with the main processor, which is located on the mother board. This error indicates that there
may be a problem with the cable connection from the front panel display or one of the main
processor ROMs may require re-seating in its socket. Check to be sure there is a proper cable
connection from the front panel display. ROMs may be reseated as follows:
1.
Turn off the power, and disconnect the line cord and all other test leads and cables from
the instrument.
2.
Remove the case cover as outlined in Section 5.
3.
Find the two firmware ROMs, U156 and U157, located on the mother board. These are
the only ICs installed in sockets. (Refer to the component layout drawing at the end of
Section 6 for exact locations.)
4.
Carefully push down on each ROM IC to make sure it is properly seated in its socket.
CAUTION
Be careful not to push down excessively, or you might crack the mother
board.
5.
Connect the line cord, and turn on the power. If the problem persists, additional
troubleshooting will be required.
4-18
Troubleshooting
Summary of Contents for 2010
Page 14: ...1 Performance Verification 1 Performance Verification...
Page 28: ...2 Calibration 2 Calibration...
Page 53: ...3 Routine Maintenance 3 Routine Maintenance...
Page 56: ...4 Troubleshooting 4 Troubleshoot ing...
Page 74: ...5 Disassembly 5 Disassembly...
Page 83: ...CONFIDENTIAL PROPRIETARY...
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ...6 Replaceable Parts 6 Replaceable Parts...
Page 98: ...CONFIDENTIAL PROPRIETARY...
Page 99: ...CONFIDENTIAL PROPRIETARY...
Page 100: ......
Page 101: ...A Specifications A Specifications...
Page 107: ...B Calibration CommandReference B Calibration CommandRef erence...