Keithley Instruments, Inc.
Cleveland, Ohio 44139
DRN.
CKD.
APP.
DATE
DATE
DATE
PART NUMBER
LTR
REVISIONS
APP. DATE
FORM 28777A-SBG
BRUNING 40-21 62198-SBG
PURCHASED ITEM
SPECIFICATIONS
2010 Low-Noise Multimeter
HW 9/5/03
Rev. C
FREQUENCY AND PERIOD CHARACTERISTICS
1,2
RESOLUTION
ACCURACY
ACV
FREQUENCY
PERIOD
GATE
±(ppm of
90 Day/1 Year
RANGE
RANGE
RANGE
TIME
reading)
±(% of reading)
100 mV
3 Hz
333 ms
to
to
to
1 s
0.3
0.01
750 V
500 kHz
2 µs
FREQUENCY NOTES
1. Specifications are for square wave inputs only. Input signal must be >10% of ACV range. If input is
<20mV on the 100mV range then the frequency must be >10Hz.
2. 20% overrange on all ranges except 750V range.
TEMPERATURE CHARACTERISTICS
THERMOCOUPLE
2, 3, 4
ACCURACY
1
90 DAY/1 YEAR (23°C ± 5°C)
RELATIVE TO SIMULATED
USING
TYPE
RANGE
RESOLUTION
REFERENCE JUNCTION
2001-TCSCAN
5
J
–200 to + 760°C
0.001°C
±0.5°C
±0.65°C
K
–200 to +1372°C
0.001°C
±0.5°C
±0.70°C
N
–200 to +1300°C
0.001°C
±0.5°C
±0.70°C
T
–200 to + 400°C
0.001°C
±0.5°C
±0.68°C
4-WIRE RTD
2, 3, 7, 8
90 DAY/1 YEAR
2 YEAR
(23°C ± 5°C)
(23°C ± 5°C)
RANGE
RESOLUTION
ACCURACY
6
ACCURACY
6
–100° to +100°C
0.001°C
±0.08°C
±0.12°C
–200° to +630°C
0.001°C
±0.14°C
±0.18°C
Temperature Notes
1. For temperatures <–100°C, add ±0.1°C and >900°C add ±0.3°C.
2. Temperature can be displayed in °C, K or °F.
3. Accuracy based on ITS-90.
4. Exclusive of thermocouple error.
5. Specifications apply to channels 2–6. Add 0.06°C/channel from channel 6.
6. Excluding probe errors.
7. 100
Ω
platinum, D100, F100, PT385, PT-3916 or user type.
8. Maximum lead resistance (each lead) to achieve rated accuracy is 5
Ω
.
INTERNAL SCANNER SPEED
MAXIMUM INTERNAL SCANNER RATES: RANGE: CHANNELS/s
1
TRIGGER DELAY = 0
2 WIRE
4 WIRE
T/C
RTD
DCV
2
ACV
2, 3
OHMS
2
OHMS
2
TEMPERATURE
2
TEMPERATURE
2
All : 105
All : 96 All : 102
<10M
Ω
: 55
All : 70
All : 2
TRIGGER DELAY = AUTO
2 WIRE
4 WIRE
T/C
RTD
DCV
2
ACV
2, 3
OHMS
2
OHMS
2
TEMPERATURE
2
TEMPERATURE
2
0.1 V : 100
All : 1.8
100
Ω
: 82
100
Ω
: 42
All : 70
All : 2
1 V : 100
1 k
Ω
: 85
1 k
Ω
: 42
10 V : 100
10 k
Ω
: 42
10 k
Ω
: 25
100 V : 70
100 k
Ω
: 28
100 k
Ω
: 21
1000 V : 70
1 M
Ω
: 8
1 M
Ω
: 7
10 M
Ω
: 5
10 M
Ω
: 5
100 M
Ω
: 3
100 M
Ω
: 3
Internal Scanner Speed Notes
1. Speeds are for 60Hz or 50Hz operation using factory default operating conditions (*RST). Auto Zero off,
Auto Range off, Display off, sample count = 1024.
2. NPLC = 0.01.
3. DETector BANDwidth: 300.
Triggering and Memory
READING HOLD SENSITIVITY:
0.01%, 0.1%, 1%, or 10% of reading.
TRIGGER DELAY:
0 to 99 hrs (1ms step size).
EXTERNAL TRIGGER DELAY:
<1ms.
EXTERNAL TRIGGER JITTER:
<500µs.
MEMORY:
1024 readings.
Math Functions
Rel, Min/Max/Average/StdDev (of stored reading), dB, dBm, Limit Test,
%, and mX+b with user defined units displayed.
dBm REFERENCE RESISTANCES:
1 to 9999
Ω
in 1
Ω
increments.
REMOTE INTERFACE
Keithley 199/196 Emulation
GPIB (IEEE-488.2) and RS-232C
SCPI (Standard Commands for Programmable Instruments)
GENERAL
POWER SUPPLY:
100V / 120V / 220V / 240V.
LINE FREQUENCY:
50Hz to 60Hz and 440Hz, automatically sensed at
power-up.
POWER CONSUMPTION:
22VA.
OPERATING ENVIRONMENT:
Specified for 0°C to 50°C. Specified to
80% R.H. at 35°C. Altitude up to 2000 meters.
STORAGE ENVIRONMENT:
–40°C to 70°C.
WARRANTY:
3 years.
SAFETY:
Conforms to European Union Directive 73/23/EEC
EN61010-1, CAT II.
EMC:
Complies with European Union Directive 89/336/EEC,
EN61326-1.
VIBRATION:
MIL-PRF-28800F Class 3 Random.
WARMUP:
2 hours to rated accuracy.
DIMENSIONS:
Rack Mounting:
89mm high
×
213mm wide
×
370mm deep (3
1
⁄
2
in
×
8
3
⁄
8
in
×
14
9
⁄
16
in).
Bench Configuration (with handle and feet):
104mm high
×
238mm wide
×
370mm deep (4
1
⁄
8
in
×
9
3
⁄
8
in
×
14
9
⁄
16
in).
SHIPPING WEIGHT:
5kg (11 lbs).
VOLT HERTZ PRODUCT:
≤
8
×
10
7
V·Hz.
ACCESSORIES SUPPLIED:
Model 1751 Safety Test Leads, User
Manual, Service Manual.
Specifications are subject to change without notice.
SPEC-2010
A C C E S S O R I E S A V A I L A B L E
1050
Padded Carrying Case with handle and should strap
1754
Universal Test Lead Kit
2000-SCAN
10-Channel Scanner
2001-TCSCAN 9-Channel Thermocouple Scanner
(includes 1-channel reference junction)
2010-EW
1-Year Warranty Extension
4288-1
Single Fixed Rack Mount Kit
4288-2
Dual Fixed Rack Mount Kit
5804
4 Terminal Test Lead Set
5805
Kelvin Probes
5806
Kelvin Clip Lead Set
5807-7
Helical Spring Point Test Leads
7007-1
Shielded GPIB Cable, 1m (3.2 ft)
7007-2
Shielded GPIB Cable, 2m (6.5 ft)
7009-5
Shielded RS-232 Cable 1.5m (5 ft)
8502
Trigger Link Adapter to 6 female BNC connector
8503
Trigger Link Cable to 2 Male BNCs, 1m (3.2 ft.)
8605
High Performance Modular Test Leads
8606
High Performance Probe Tip Kit
A
18127 Released
SZ
1/10/96
B
27373 Revised
SZ
2/21/03
C
28238 Revised
SZ
9/5/03
CAB
1/9/96
DMS
1/9/96
Summary of Contents for 2010
Page 14: ...1 Performance Verification 1 Performance Verification...
Page 28: ...2 Calibration 2 Calibration...
Page 53: ...3 Routine Maintenance 3 Routine Maintenance...
Page 56: ...4 Troubleshooting 4 Troubleshoot ing...
Page 74: ...5 Disassembly 5 Disassembly...
Page 83: ...CONFIDENTIAL PROPRIETARY...
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ...6 Replaceable Parts 6 Replaceable Parts...
Page 98: ...CONFIDENTIAL PROPRIETARY...
Page 99: ...CONFIDENTIAL PROPRIETARY...
Page 100: ......
Page 101: ...A Specifications A Specifications...
Page 107: ...B Calibration CommandReference B Calibration CommandRef erence...