AC voltage calibration
Follow these steps for AC voltage calibration:
1.
Connect the calibrator to the Model 2010 INPUT HI and LO terminals as shown in
2.
Perform the calibration steps summarized in
. For each step:
•
Set the calibrator to the indicated value, and make sure the calibrator is in operate.
•
Press ENTER to complete each step.
•
Wait until the Model 2010 completes each step.
NOTE
Do not exceed ±15V peak between sense LO and LO. You may have to remove the
external sense leads before calibrating AC volts. Even though some calibrators can
turn off their external sense, the AC voltage can still cause excess voltage to appear
between sense LO and LO.
Table 2-6
AC voltage calibration summary
Calibration step
Calibrator voltage, frequency
10mV AC at 1kHz
100mV AC at 1kHz
100mV AC at 50kHz
1V AC at 1kHz
1V AC at 50kHz
10V AC at 1kHz
10V AC at 50kHz
100V AC at 1kHz
100V AC at 50kHz
700V AC at 1kHz
10.00000mV, 1kHz
100.0000mV, 1kHz
100.0000mV, 50kHz
1.000000V, 1kHz
1.000000V, 50kHz
10.00000V, 1kHz
10.00000V, 50kHz
100.0000V, 1kHz
100.0000V, 50kHz
700.000V, 1kHz
2000 MULTIMETER
!
R
Output HI
Input
LO
Output
LO
Model 2010
5700A Calibrator
Input HI
Figure 2-4
Connections for AC
volts calibration
Calibration
2-11
Summary of Contents for 2010
Page 14: ...1 Performance Verification 1 Performance Verification...
Page 28: ...2 Calibration 2 Calibration...
Page 53: ...3 Routine Maintenance 3 Routine Maintenance...
Page 56: ...4 Troubleshooting 4 Troubleshoot ing...
Page 74: ...5 Disassembly 5 Disassembly...
Page 83: ...CONFIDENTIAL PROPRIETARY...
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ...6 Replaceable Parts 6 Replaceable Parts...
Page 98: ...CONFIDENTIAL PROPRIETARY...
Page 99: ...CONFIDENTIAL PROPRIETARY...
Page 100: ......
Page 101: ...A Specifications A Specifications...
Page 107: ...B Calibration CommandReference B Calibration CommandRef erence...