:DATE
(:CALibration:PROTected:DATE)
Purpose
To program the calibration date.
Format
:cal:prot:date <year>, <month>, <day>
Parameters
<year> = 1997 to 2096
<month> = 1 to 12
<day> = 1 to 31
Query
:cal:prot:date?
Response
<year>, <month>, <day>
Description
The
:DATE
command allows you to store the calibration date in instrument
EEROM for future reference. You can read back the date from the instrument
by using the
:DATE?
query.
NOTE
The year, month, and day parameters must be delimited by commas.
Example
:CAL:PROT:DATE 1997,11,20
Send cal date (11/20/97).
:NDUE
(:CALibration:PROTected:NDUE)
Purpose
To send the next calibration due date to the instrument.
Format
:cal:prot:ndue <year>, <month>, <day>
Parameters
<year> = 1997 to 2096
<month> = 1 to 12
<day> = 1 to 31
Query
:cal:prot:ndue?
Response
<year>, <month>, <day>
Description
The
:NDUE
command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date by using the
:NDUE?
query.
NOTE
The next due date parameters must be delimited by commas.
Example
:CAL:PROT:NDUE 1998,11,20
Send due date (11/20/98).
B-8
Calibration Reference
Summary of Contents for 2182
Page 10: ......
Page 14: ...1 Performance Verification Performance Verification...
Page 25: ...1 12 Performance Verification...
Page 26: ...2 Calibration Calibration...
Page 43: ...2 18 Calibration...
Page 44: ...3 Routine Maintenance RoutineMain tenance...
Page 47: ...3 4 Routine Maintenance...
Page 48: ...4 Troubleshooting Troubleshoot ing...
Page 64: ...5 Disassembly Disassembly...
Page 73: ...5 10 Disassembly...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ...6 Replaceable Parts Replaceable Parts...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ...A Specifications Specifica tions...
Page 96: ...B CalibrationReference Calibration Reference...
Page 109: ...B 14 Calibration Reference...
Page 110: ...C CalibrationProgram Calibration Program...
Page 115: ...C 6 Calibration Program...
Page 119: ......