Introduction
This appendix includes a calibration program written in BASIC to help you to field calibrate
the Model 2182. Refer to
for more details on calibration procedures, equipment, and
connections.
covers calibration commands in detail.
Computer hardware requirements
The following computer hardware is required to run the calibration programs:
•
IBM PC compatible computer.
•
Keithley KPC-488.2, KPS-488.2, or KPC-488.2AT, or CEC PC-488 IEEE-488 interface
for the computer.
•
Two shielded IEEE-488 connecting cables (Keithley Model 7007).
Software requirements
In order to use the calibration programs, you will need the following computer software:
•
Microsoft QBasic (supplied with MS-DOS 5.0 or later).
•
MS-DOS version 5.0 or later.
•
HP-style Universal Language Driver, CECHP.EXE (supplied with Keithley and CEC
interface cards listed above).
Calibration equipment
The following calibration equipment is required:
•
Fluke 5700A calibrator
•
Low thermal input cable
•
Low thermal shorting plug
for detailed equipment specifications.
C-2
Calibration Program
Summary of Contents for 2182
Page 10: ......
Page 14: ...1 Performance Verification Performance Verification...
Page 25: ...1 12 Performance Verification...
Page 26: ...2 Calibration Calibration...
Page 43: ...2 18 Calibration...
Page 44: ...3 Routine Maintenance RoutineMain tenance...
Page 47: ...3 4 Routine Maintenance...
Page 48: ...4 Troubleshooting Troubleshoot ing...
Page 64: ...5 Disassembly Disassembly...
Page 73: ...5 10 Disassembly...
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ...6 Replaceable Parts Replaceable Parts...
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ...A Specifications Specifica tions...
Page 96: ...B CalibrationReference Calibration Reference...
Page 109: ...B 14 Calibration Reference...
Page 110: ...C CalibrationProgram Calibration Program...
Page 115: ...C 6 Calibration Program...
Page 119: ......