5-16
Ratio and Delta
SCPI programming - ratio and delta
Programming examples
Ratio programming example
— The following program fragment enables Ratio and
displays the result on the computer CRT.
CALL SEND (7, “:sens:volt:func ‘volt’”, status%) ‘ Select voltage function.
CALL SEND (7, “:sens:volt:chan 1”, status%)
‘ Select Channel 1 (DCV1) for
‘ range control.
CALL SEND (7, “:sens:volt:ratio”, status%)
‘ Enable Ratio.
CALL SEND (7, “:sens:data:fresh?”, status%)
‘ Request a fresh reading.
reading$ = SPACE$(80)
CALL ENTER(reading$, length%, 7, status%)
‘ Address 2182 to talk.
PRINT reading$
‘ Display Ratio reading on CRT.
Table 5-1
SCPI commands - ratio and delta
Commands
Description
Default
:SENSe[1]
SENSe Subsystem:
:FUNCtion <name>
Select voltage function; ‘VOLTage’.
VOLT
:CHANnel <chan>
Select range control channel; 1 or 2.
:VOLTage[:DC]
Path to configure DC volts:
:RATio <b>
Enable or disable Ratio (V1/V2).
OFF
:DELTa <b>
Enable or disable Delta. Not valid with
TEMP1 or TEMP2 selected.
OFF
:SYSTem:FAZero[:STATe] <b>
Enable or disable Front Autozero. To double
the speed of Delta, disable Front Autozero.
ON
Note: Enabling Ratio disables Delta and conversely, enabling Delta disables Ratio.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...