SCPI programming - limits
For remote operation, the testing capabilities of Limit 1 and Limit 2 are the same. Limit 1
and/or Limit 2 can be enabled. The commands to configure and control limit testing are listed in
NOTE
When testing limits remotely, keep in mind that the front panel HI/IN/LO status
messages only apply to Limit 1. Also, if the front panel beeper is set for OUTSIDE or
INSIDE, it will operate according to its front panel definition as previously explained.
Table 8-1
SCPI commands - limits
Commands
Description
Default
Note
:CALCulate3
CALCulate3 Subsystem:
:LIMit
Configure and control Limit 1:
:UPPer <n>
Set upper (HI1) limit; -100e6 to 100e6.
1
:LOWer <n>
Set lower (LO1) limit; -100e6 to 100e6.
–1
:STATe <b>
Enable or disable Limit 1 test.
OFF
:FAIL?
Query test result; 0 = pass, 1 = fail.
1
:CLEar
Path to clear fail indication:
2
[:IMMediate]
Clear fail indication.
:AUTO <b>
Enable or disable auto clear.
ON
:LIMit2
Configure and control Limit 2:
:UPPer <n>
Set upper (HI2) limit; -100e6 to 100e6.
2
:LOWer <n>
Set lower (LO2) limit; -100e6 to 100e6.
–2
:STATe <b>
Enable or disable Limit 2 test.
OFF
:FAIL?
Query test result; 0 = pass, 1 = fail.
1
:CLEar
Path to clear fail indication:
2
[:IMMediate]
Clear fail indication.
:AUTO <b>
Enable or disable auto clear.
ON
:IMMediate
Re-perform limit tests.
3
Limits
8-5
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...