9-8
Stepping and Scanning
External Stepping/Scanning
The settings for external stepping/scanning are explained as follows:
Min/Max Values
— These two values specify the beginning and ending channels for the
step/scan list. Valid values for Min is 1 to 799, and valid values for Max is 2 to 800. However,
the Max value must be larger than the Min value.
Timer
— The maximum timer interval is 99H:99M:99.999S (Hour:Minute:Second format).
Reading Count
— This indicates the total number of measurements that will be performed
for each step/scan cycle. For example, if Min is set to 1 and Max is set to 10, the Reading Count
will automatically set to 10. For each additional step/scan cycle, simply add 10 to the Reading
Count. Therefore, to perform three step/scan cycles, set the Reading Count to 30. Reading Count
can be set from 2 to 1024.
Perform the following steps to configure external stepping or scanning:
1.
Press SHIFT and then CONFIG. Use the
key to display the present SCANNING type
(INTernal or EXTernal).
2.
Press the
or
key to display EXT, and press ENTER.
3.
The minimum channel (MIN CHAN) to step/scan is displayed. Use the edit keys (
,
,
, and
) to specify the Min channel, and press ENTER.
4.
The maximum channel (MAX CHAN) is then displayed. Use the edit keys to specify the
Max channel, and press ENTER.
5.
The present state of the timer is displayed (OFF or ON). Press
or
to display the
desired timer state, and press ENTER.
6.
If you turned the timer on, the timer interval will be displayed. Use the edit keys (
,
,
, and
) to set the timer interval, and press ENTER.
7.
The Present Reading Count (RNG CNT) is displayed. It will be (Max - Min) + 1. If you
wish to increase the Reading Count, use the edit keys to display the value, and press
ENTER to return to the normal display state.
Stepping/Scanning examples
Internal scanning
S
ettings:
Control Source = Immediate (timer off)
Delay = Auto
Channel 1 Count = 4
Reading Count = 10
Overview:
With Channel 1 count set to 4, each scan cycle will measure Channel 2
once and Channel 1 four times, for a total of five measurements. This
sets the Sample Counter in
to 5. The Reading Count 10
indicates that the five measurement scan will be performed twice. This
sets the Trigger Counter to 2. A total of 10 measurements will be
performed. All ten readings will be stored in the buffer.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...