Delta, Pulse Delta, and Differential Conductance
I-5
Test system configurations
There are two test system configurations that can be used for Delta, Pulse Delta, and
Differential Conductance measurements and are shown in
panel stand-alone operation and the other is for remote programming (PC control system).
Both systems use serial communications (via RS-232 interface) between the Model 622x
and the Model 2182/2182A. The Model 622x sends setup commands to the
Model 2182/2182A, and the Model 2182/2182A sends Delta, Pulse Delta, or Differential
Conductance readings to the buffer of the Model 622x. Once the test is started, trigger
synchronization between the two instruments is controlled by the Trigger Link.
Figure I-2
Test system configurations
Trigger Link
Keithley
622x
Keithley
2182/2182A
Current Source
Nanovoltmeter
RS-232
A) Stand-alone system (front panel operation)
Trigger Link
Current Source
Nanovoltmeter
RS-232
B) PC control of 6220/21
PC
(null-modem)
(null-modem)
IEEE-488
or
Ethernet (6221)
RS-232 On
Keithley
2182/2182A
RS-232 On
GPIB or
Ethernet (6221)
Selected
Keithley
622x
GPIB or
Ethernet (6221)
Selected
DUT
Out
Hi
Lo
Ch 1
Hi
Lo
DUT
Out
Hi
Lo
Ch 1
Hi
Lo
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...