5-24
Ratio and Delta
To check measurement repeatability, you may wish to perform more than one Delta
measurement at each current amplitude. In
, the SourceMeter outputs five bipolar
steps for each amplitude. The result will be five Delta measurements for each amplitude. When
configuring the custom sweep, ±10µA would be assigned to the first 10 points of the sweep,
±20µA would be assigned to the next 10 points, and ±50µA would be assigned to the last 10
points. Therefore, the custom sweep in
would be made up of 30 points (P0 through
P29).
The procedure to use the SourceMeter and Model 2182 to perform Delta measurements is
provided in “
Delta measurement procedure using a SourceMeter
.” That procedure (presented
earlier in this section under “
”) uses a 2-point custom sweep and will have to be modified
for this application as follows:
•
This application uses two Model 2182s. Therefore, both nanovoltmeters must be
configured exactly the same.
•
In Step 3 of the procedure, change the Trigger Count to equal the number of points in the
custom sweep. For example, if using the 30-point custom sweep in
, set the
Trigger Count to 30.
•
In Step 5, assign current values to the sweep points. For the example 30-point sweep, the
current values for points P0 through P29 are shown in
.
•
In Step 7, enable line synchronization on both Model 2182s. To access control, press
SHIFT and then LSYNC.
NOTE
Optimum synchronization among all instruments is achieved when Model 2182 line
synchronization is enabled, and autozero is disabled. Autozero cannot be disabled
from the front panel of the Model 2182. When controlling this application over the
bus, use the following commands for the Model 2182s:
:SYSTem:LSYNc:STATe ON
Enable 2182 Line Synchronization
:SYSTem:AZERo:STATe OFF
Disable 2182 Autozero
Autozero should only be disabled for short periods of time. After performing a sweep,
re-enable Autozero. For details on Autozero, see
.
•
In Step 9, set the buffers of both Model 2182s to store the Delta readings. If performing
a 30-point sweep, set both buffers to store 30 Delta readings.
Log sweep
— If your test requires currents that span two or more decades, you can configure
the SourceMeter to output a log sweep. The 30-point sweep in
is confined to the
10µA-100uA decade. If, for example, you want to expand the sweep to span three current
decades, the next 30 sweep points would be used for the 100µA, 200µA, and the 500µA
amplitudes. The last 30 sweep points would be used for the 1mA, 2mA, and 5mA amplitudes.
The I-V data points can then be graphed using a logarithmic/linear scale.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...