External triggering with BNC connections
An adapter cable is available to connect the micro-DIN Trigger Link of the Model 2182 to
instruments with BNC trigger connections. The Model 8503 DIN to BNC Trigger Cable has a
micro-DIN connector at one end and two BNC connectors at the other end. The BNC cables are
labeled VMC (trigger line 1) and EXT TRIG (trigger line 2).
shows how a Keithley Model 220 Current Source can be connected to the Trigger
Link of the Model 2182 using the adapter cable. When used with the STEP mode of the
Model 220, you can perform synchronized source-measure operations without the use of a
computer. Whenever the Model 220 receives a trigger from the Model 2182, it will step to the
next current source value.
Figure 7-9
DIN to BNC trigger cable
NOTE
If the Model 2182 trigger line configuration has been changed from the factory
setting, the Model 8502 Trigger Link Adapter must be used to interface with
instruments having BNC trigger connections. It has two micro-DIN connectors and
six BNC connectors, one for each trigger line.
2182 Nanovoltmeter
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
RS232
120
1
3
5
2
4
6
VMC
EXT TRIG
FUSE
LINE
250mAT
(SB)
100 VAC
120 VAC
125mAT
(SB)
220 VAC
240 VAC
WITH ANALOG OUTPUT
GAIN SET TO 1:
±
FULL SCALE READINGS
PRODUCE
±
1V OUTPUT
TRIGGER
LINK
!
LINE RATING
50, 60
400HZ
17 VA MAX
IEEE-488
(CHANGE IEEE ADDRESS
FROM FRONT PANEL)
!
!
ANALOG OUTPUT
1K
Ω
OUTPUT
RESISTANCE
MADE IN
U.S.A.
8503 DIN to BNC
Trigger Cable
Input
Output
External
Trigger
220 Current Source
Tigger
Link
7-12
Triggering
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...