12-8
Common Commands
*OPC – Operation Complete
Set the OPC bit in the standard event register after
all pending commands are complete
Description
After the *OPC command is sent, the Operation Complete bit (bit B0) of the Standard Event
Status Register will set immediately after the last pending command is completed. If the
corresponding bit (bit B0) in the Standard Event Enable Register and Bit 5 (Event Summary Bit)
of the Service Request Enable Register is set, the RQS/MSS (Request for Service/Master
Summary Status) bit in the Status Byte Register will set.
When used with the immediate initiation command (:INITiate), the OPC bit in the Standard
Event Status Register will not set until the Model 2182 goes back into the idle state. The :INIT
command operation is not considered finished until the Model 2182 goes back into the idle state.
See the description for *WAI for more information on command execution.
When used with the *TRG command, the OPC bit will not set until the operations associated
with the *TRG command (and the initiate command) are finished. The *TRG command is
considered to be finished when the Device Action completes or when operation stops a control
source to wait for an event (see “
,” in
To use the *OPC exclusively with the *TRG command, first force the completion of the
initiate command so that only the *TRG command is pending. Do this by sending the :ABORt
command to place the instrument in idle, which (by definition) completes the initiate command.
Since continuous initiation is on, operation continues into the Trigger Model. After sending the
*TRG command, the OPC bit sets when the *TRG command is finished.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...