Additional SCPI Commands
15-19
:VERSion? command
:VERSion?
:SYSTem:VERSion?
Read SCPI version
Description
This query command is used to read the version of the SCPI standard being used by the
Model 2182. Example code:
1991.0
The above response message indicates the version of the SCPI standard.
:ERRor? command
:ERRor?
:SYSTem:ERRor?
Read Error Queue
Description
As error and status messages occur, they are placed in the Error Queue. This query command
is used to read those messages. The Error Queue is a first-in, first-out (FIFO) register that can
hold up to ten messages. Each time you read the queue, the “oldest” message is read, and that
message is then removed from the queue.
If the queue becomes full, the “350, Queue Overflow” message occupies the last memory
location in the register. On power-up, the queue is empty. When the Error Queue is empty, the
“0, No error” message is placed in the Error Queue.
The messages in the queue are preceded by a number. Negative (–) numbers are used for SCPI
defined messages, and positive (+) numbers are used for Keithley defined messages.
lists the messages.
NOTE
The :SYSTem:ERRor? query command performs the same function as the
:STATus:QUEue? query command (see STATus subsystem).
:CLEar command
:CLEar
:SYSTem:CLEar
Clear Error Queue
Description
This action command is used to clear the Error Queue of messages.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...