Appendix D: Convert Model 2400 to Model 2461 code
Model 2461 Interactive SourceMeter® Instrument Reference Manual
D-12
2461-901-01 A/November 2015
Model 2400 command
:FORMat[:DATA]
:FORMat[:DATA]?
Model 2461 command
(on page 6-46)
:FORMat[:DATA]?
Notes
Model 2400 command
:FORMat:ELEMents[:SENSe[1]]
:FORMat:ELEMents[:SENSe[1]]?
Model 2461 command
Not available
Notes
In the Model 2461, format elements are specified as part of the
READ?
,
FETCh?
,
MEASure?
, and
TRACe:DATA?
commands with each use of the
command instead of using a global setting for all commands. The elements
may be unique for each command and are unique each time the command is
processed.
Model 2400 command
:FORMat:ELEMents:CALCulate
:FORMat:ELEMents:CALCulate?
Model 2461 command
Not available
Notes
Model 2400 command
:FORMat:SOURce2
:FORMat:SOURce2?
Model 2461 command
Not available
Notes
Model 2400 command
:FORMat:SREGister
:FORMat:SREGister?
Model 2461 command
Not available
Notes
MEASure:<function>?
Model 2400 command
:MEASure:CURRent[:DC]?
:MEASure:RESistance?
:MEASure:VOLTage[:DC]?
Model 2461 command
(on page 6-4), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
.
Notes
Can specify a buffer in Model 2461 and which buffer elements to access from
the buffer when you specify the command. In the Model 2461, this command
changes the measurement function to the function in the command if it is not
already active, makes readings, and stores them in a reading buffer, which you
can specify. When it changes to that function, it recalls the settings as they
were the last time that function was active. It does not go to factory default
settings for the function (which the Series 2400 did). Also, in the Model 2461,
this command does not map to
CONFigure
,
READ?
, and
FETCh?
.
OUTPut subsystem
Model 2400 command
:OUTPut[1]:ENABle[:STATe]
:OUTPut[1]:ENABle[:STATe]?
Model 2461 command
Not available
Notes