Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-61
Example 1
CURR:NPLC 0.5
Sets the measurement time for current
measurements to 0.0083 s (0.5/60).
Example 2
RES:NPLC 0.5
Sets the measurement time for resistance
measurements to 0.0083 s (0.5/60).
Example 3
VOLT:NPLC 0.5
Sets the measurement time for voltage
measurements to 0.0083 s (0.5/60).
Also see
Using NPLCs to adjust speed and accuracy
(on page 4-9)
[:SENSe[1]]:<function>:OCOMpensated
This command enables or disables offset compensation.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
OFF (0)
Usage
[:SENSe[1]]:<function>:OCOMpensated <state>
[:SENSe[1]]:<function>:OCOMpensated?
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
<state>
Disable offset compensation:
OFF
or
0
Enable offset compensation:
ON
or
1
Details
The voltage offsets caused by the presence of thermoelectric EMFs (V
EMF
) can adversely affect
resistance measurement accuracy. To overcome these offset voltages, you can use offset-
compensated ohms.
This feature is only applied to resistance measurements or when
[:SENSe[1]]:CURRent:UNIT
or
[:SENSe[1]]:VOLTage:UNIT
is set to
OHM
.