Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-131
smu.interlock.tripped
This attribute indicates that the interlock has been tripped.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
interlockStatus
= smu.interlock.tripped
interlockStatus
The status of the interlock:
•
smu.OFF
: The interlock is not asserted and the V range is disabled; lower
voltage ranges are available
•
smu.ON
: The interlock signal is asserted and all voltage ranges are available
Details
This command gives you the status of the interlock. When the safety interlock signal is asserted, all
voltage ranges of the instrument are available. However, when the safety interlock signal is not
asserted, the V range is disabled, limiting the nominal output to less than ±42 V.
When the interlock is not asserted:
•
The front-panel INTERLOCK indicator is off.
•
High voltage ranges are disabled.
•
If you attempt to turn on the source with a voltage more than ±21 V, an event message is generated.
Example
print(smu.interlock.tripped)
If the interlock is not asserted, returns
smu.OFF
.
If the interlock is asserted, returns
smu.ON
.
Also see
None
smu.measure.autorange
This attribute determines if the measurement range is set manually or automatically for the selected measure
function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.ON
Usage
autoRange
= smu.measure.autorange
smu.measure.autorange =
autoRange
autoRange
Set the measurement range manually:
smu.OFF
Set the measurement range automatically:
smu.ON