Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-188
2461-901-01 A/November 2015
smu.source.output
This attribute enables or disables the source output.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
smu.OFF
Usage
sourceOutput
= smu.source.output
smu.source.output =
sourceOutput
sourceOutput
Switch the source output off:
smu.OFF
Switch the source output on:
smu.ON
Details
When the output is switched on, the instrument sources either voltage or current, as set by
smu.source.func
.
Example
smu.source.output = smu.ON
Switch the source output of the instrument to
on.
Also see
Turn the Model 2461 output on or off
(on page 8-184)
(on page 8-186)
smu.source.protect.level
This attribute sets the overvoltage protection setting of the source output.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
smu.PROTECT_NONE
Usage
limit
= smu.source.protect.level
smu.source.protect.level =
limit
limit
The overvoltage protection value; set as
smu.PROTECT_
x
, where
x
is
2V
,
5V
,
10V
,
20V
,
40V
,
60V
,
80V
, or
NONE