Command errors include:
•
IEEE Std 488.2 syntax error:
The instrument received a message that does not follow the defined
syntax of the IEEE Std 488.2 standard.
•
Semantic error:
The instrument received a command that was misspelled or received an optional IEEE
Std 488.2 command that is not implemented in the instrument.
•
GET error:
The instrument received a Group Execute Trigger (GET) inside a program message.
Example 1
standardRegister = status.standard.OPC + status.standard.QYE
status.standard.enable = standardRegister
Uses constants to set the OPC and QYE bits of the standard event status enable register.
Example 2
-- decimal 5 = binary 0000 0101
standardRegister = 5
status.standard.enable = standardRegister
Uses a decimal value to set the OPC and
QYE bits of the standard event status
enable register.
Also see
(on page C-3)
(on page C-16)
status.standard.event
This attribute returns the contents of the Standard Event Status Register set of the status model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
status.preset()
Not applicable
0
Usage
standardRegister
= status.standard.event
standardRegister
The status of the standard event status register