Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 3: Functions and features
2461-901-01 A/November 2015
3-39
Storing configuration index 2
Refer to the instructions in
(on page 3-38), and continue as follows to
configure the instrument for configuration index 2.
Using the front panel to configure the instrument:
Change the following instrument settings to configure the instrument with the settings you want to
save for configuration index 2:
•
Set the measure range to 100 mA
•
Set NPLC to 2
Using the front panel to store the active settings to configuration index 2:
1. Return to the configuration list menu.
2. Select
MyMeasList
.
3. Select
System to Index
. This saves the active system settings to the configuration index.
Select
View Details
to see some of the settings stored in configuration index 2. See
settings stored in a measure configuration list
(on page 3-33) for information about the settings
that are stored.
4. Select
HOME
to return to the Home screen.
The procedure above is for a measure configuration list. If you want to create a source configuration
list, under
Source
, select
Config List
.
Recalling a configuration index
You can recall the settings stored in a specific configuration index in a configuration list.
For example, use the following procedure to recall configuration index 2 from
MyMeasList
.
Recall source configuration lists before measure configuration lists. This order ensures that
dependencies between source and measure settings will be properly handled.
Using the front panel to recall a configuration index:
1. Press the
MENU
key.
2. Under Measure, select
Config Lists
. The MEASURE CONFIGURATION LISTS screen is
displayed.
If you want to recall an index on a source configuration list, under
Source
, select
Config List
.
3. Choose
Select List
. A menu of available configuration lists is displayed.
4. Select
MyMeasList
. The configuration indexes in the list display.
5. Select the second configuration index.
6. Select
Index to System
.