Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2461-901-01 A/November 2015
6-33
Example
:CALC2:VOLT:LIM1:CLE:AUTO OFF
:CALC2:VOLT:LIM1:AUD FAIL
:CALC2:VOLT:LIM1:LOW 0.25
:CALC2:VOLT:LIM1:UPP 2.5
:CALC2:VOLT:LIMIT1:STAT ON
:READ?
:CALC2:VOLT:LIMIT1:FAIL?
:CALC2:VOLT:LIM1:CLE
Set limit autoclear off.
Enable the beeper for limit 1 when a voltage
measurement exceeds the limit.
Set lower limit 1 for voltage to 0.25 V.
Set upper limit 1 for voltage to 2.5 V.
Enable limit 1 testing for voltage.
Make a reading; the limit is checked and results
display on the front panel
Return the test results; example output if the test
fails on the low limit:
LOW
Clear the test results.
Also see
:CALCulate2:<function>:LIMit<Y>:CLEar:AUTO
(on page 6-26)
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
(on page 6-28)
:CALCulate2:<function>:LIMit<Y>:FAIL?
(on page 6-29)
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]
(on page 6-30)
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]
(on page 6-33)
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]
This command specifies the upper limit for a limit test.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
1
Usage
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA] <n>
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA] DEFault
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA] MINimum
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA] MAXimum
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]?
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? DEFault
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? MINimum
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? MAXimum
<function>
The measure function:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
The digitize function:
•
Digitize current:
DIGitize:CURRent
•
Digitize voltage:
DIGitize:VOLTage
<Y>
Limit number:
1
or
2
<n>
The value of the upper limit (
−
9.11 to +9.11)