Section 6: SCPI command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
6-132
2461-901-01 A/November 2015
Example
*RST
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SENS:FUNC "CURR"
SENS:CURR:RANG 100e-6
SOUR:SWE:VOLT:LOG .1, 10, 20, 1e-3, 1, FIXED
INIT
Reset the instrument to its defaults.
Set the source function to voltage.
Set the source range to 20 V.
Set the measure function to current.
Set the current range to 100
µ
A.
Set up a log sweep that sweeps from 0.1 to 10 V in 20 steps with a source delay of 1 ms, a sweep count of 1,
and a fixed source range.
Start the sweep.
Also see
(on page 6-182)