Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-175
smu.reset()
This function turns off the output and resets the commands that begin with
smu.
to their default settings.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu.reset()
Details
This function turns off the output and resets the commands that begin with
smu.
to their default
settings.
Example
smu.reset()
Turns off the output and resets the SMU
commands to their default settings.
Also see
(on page 8-97)
smu.source.autorange
This attribute determines if the range is selected manually or automatically for the selected source function or
voltage source.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Source configuration list
Function change
Configuration script
Source configuration list
smu.ON
Usage
sourceAutorange
= smu.source.autorange
smu.source.autorange =
sourceAutorange
sourceAutorange
Disable automatic source range:
smu.OFF
Enables automatic source range:
smu.ON
Details
This command indicates the state of the range for the selected source. When automatic source range
is disabled, the source range is set manually.
When automatic source range is enabled, the instrument selects the range that is most appropriate
for the value that is being sourced. The output level controls the range. If you read the range after the
output level is set, the instrument returns the range that the instrument chose as appropriate for that
source level.
If the source range is set to a specific value from the front panel or a remote command, the setting for
automatic range is set to disabled.
Only available for the current and voltage functions.