Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-194
2461-901-01 A/November 2015
smu.source.pulsesweeplinear()
This function sets up a linear pulse sweep for a fixed number of pulse points.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, MeasEnable, bufferName, sDelay, offTime, count
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, MeasEnable, bufferName, sDelay, offTime, count, xBiasLimit
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort
)
smu.source.pulsesweeplinear(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort, dual
)
configListName
The name of the source configuration list that the sweep uses; this must be defined
before sending this command
biasLevel
Output level the instrument sources before the first pulse and returns to between
pulses (delay and off time):
•
Current: -7.35 A to 7.35 A
•
Voltage: -105 V to 105 V
start
The voltage or current source level at which the pulse sweep starts:
•
Current: -10.5 A to 10.5 A
•
Voltage: -105 V to 105 V
stop
The voltage or current source level at which the pulse sweep stops:
•
Current: -10.5 A to 10.5 A
•
Voltage: -105 V to 105 V
points
The number of pulse-measure points between the start and stop values of the pulse
sweep (2 to 1e6)
pulseWidth
The time at the amplitude level for each pulse:
•
Extended operating area: 10 A at 100 V, 5 % duty cycle, maximum pulse
width: 1 ms, minimum pulse width 150 µs (load dependent)
•
Normal operating area (DC): 99.99 % duty cycle, maximum pulse width
10,000 s, minimum pulse width 150 µs
measEnable
Enable or disable measurements at the top of each pulse:
•
Enable:
smu.ON
(default)
•
Disable:
smu.OFF