Section 8: TSP command reference
Model 2461 Interactive SourceMeter® Instrument Reference Manual
8-202
2461-901-01 A/November 2015
Also see
(on page 3-30)
(on page 3-64)
(on page 8-214)
(on page 8-259)
(on page 3-107)
smu.source.pulsesweeplog()
This function sets up a logarithmic pulse sweep for a set number of source points.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort, dual
)
smu.source.pulsesweeplog(
configListName, biasLevel, start, stop, points,
pulseWidth, measEnable, bufferName, sDelay, offTime, count, xBiasLimit,
xPulseLimit, failAbort, dual, asymptote
)
configListName
The name of the source configuration list that the sweep uses; this must be defined
before sending this command
biasLevel
Output level the instrument sources before the first pulse and returns to between
pulses (delay and off time):
•
Current: -7.35 A to 7.35 A
•
Voltage: -105 V to 105 V
start
The voltage or current source level at which the pulse sweep starts:
•
Current: -10.5 A to 10.5 A
•
Voltage: -105 V to 105 V