Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-221
Details
This command selects which set of input and output terminals the instrument uses. You can select
front panel or rear panel terminals.
If the output is turned on when you change from one set of terminals to the other, the output is turned
off.
This command replaces the
smu.measure.terminals
command, which is deprecated.
Example
smu.terminals = smu.TERMINALS_FRONT
Use the front-panel terminals for measurements.
Also see
None
status.clear()
This function clears event registers and the event log.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
status.clear()
Details
This command clears the event registers of the Questionable Event and Operation Event Register
set. It does not affect the Questionable Event Enable or Operation Event Enable registers.
Example
status.clear()
Clear the bits in the registers
Also see
(on page B-2)
status.condition
This attribute stores the status byte condition register.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
statusByte
= status.condition
statusByte
The status byte