Model 2461 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2461-901-01 A/November 2015
8-303
trigger.timer[N].delay
This attribute sets and reads the timer delay.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger timer
N
reset
Configuration script
10e-6 (10
µ
s)
Usage
interval
= trigger.timer[
N
].delay
trigger.timer[
N
].delay =
interval
interval
Delay interval in seconds (8 µs to 100 ks)
N
Trigger timer number (1 to 4)
Details
Once the timer is enabled, each time the timer is triggered, it uses this delay period.
Assigning a value to this attribute is equivalent to:
trigger.timer[
N
].delaylist = {
interval
}
This creates a delay list of one value.
Reading this attribute returns the delay interval that will be used the next time the timer is triggered.
If you use the trigger timer with a trigger model, make sure the trigger timer delay is set so that the
readings are paced correctly.
Example
trigger.timer[1].delay = 50e-6
Set the trigger timer 1 to delay for
50 µs.
Also see
(on page 8-305)
trigger.timer[N].delaylist
This attribute sets an array of timer intervals.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger timer
N
reset
Configuration script
{10e-6}
Usage
intervals
= trigger.timer[
N
].delaylist
trigger.timer[
N
].delaylist =
intervals
intervals
Table of delay intervals in seconds
N
Trigger timer number (1 to 4)