B-4
Calibration Reference
:LOCK
(:CALibration:PROTected:LOCK)
Purpose
To lock out calibration.
Format
:cal:prot:lock
Query
:cal:prot:lock?
Response
0
Calibration unlocked
1
Calibration locked
Description
The :LOCK command allows you to lock out calibration after complet-
ing the procedure. Thus, :LOCK performs the opposite of sending the
password with the :CODE command. The :LOCK? query returns cali-
bration lock status.
Note
To unlock calibration, send the :CODE command with the appropriate
password.
Example
:CAL:PROT:LOCK
Lock out calibration.
:SAVE
(:CALibration:PROTected:SAVE)
Purpose
To save calibration constants in EEROM after the calibration procedure.
Format
:cal:prot:save
Description
The :SAVE command stores internally calculated calibration constants
derived during calibration in EEROM. EEROM is non-volatile memory,
and calibration constants will be retained indefinitely once saved. Gen-
erally, :SAVE is sent after all other calibration steps (except for
:LOCK).
Note
Calibration will be only temporary unless the :SAVE command is sent to per-
manently store calibration constants. Calibration data will not be saved if cal-
ibration was not unlocked by sending the :CODE command or if invalid
calibration data exists.
Example
:CAL:PROT:SAVE
Save calibration constants.
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...