Calibration
2-15
Table 2-5
Current calibration values
Current range
Calibration currents
2nA
+1.90000nA
0.0000nA
-1.90000nA
20nA
+19.0000nA
0.0000nA
-19.0000nA
200nA
+190.000nA
0.000nA
-190.000nA
2
µ
A
+1.90000
µ
A
0.00000
µ
A
-1.90000
µ
A
20
µ
A
+19.0000
µ
A
0.0000
µ
A
-19.0000
µ
A
2
00µ
A
+190.000
µ
A
0.000
µ
A
-190.000
µ
A
2mA
+1.90000mA
0.00000mA
-1.90000mA
20mA
+19.0000mA
0.0000mA
-19.0000mA