2-20
Calibration
NOTE
For a detailed description of all calibration commands and queries, refer to
Appendix B.
Remote calibration procedure
Step 1. Prepare the Model 2500 for calibration
1.
With the power off, connect the Model 2500 to the controller IEEE-488 interface or
RS-232 port using a shielded interface cable.
2.
Turn on the Model 2500 and the test equipment, and allow them to warm up for at
least one hour before performing calibration.
3.
If you are using the IEEE-488 interface, make sure the primary address of the
Model 2500 is the same as the address specified in the program you will be using to
send commands. (Use the MENU key and the COMMUNICATION menu to
access the IEEE-488 address.)
Table 2-6
Remote calibration command summary
Command
Description
:CALibration
:PROTected
:CODE ‘<password>’
:SENSe[1] <NRf>
:SENSe2 <NRf>
:SOURce[1] <NRf>
:OFFSet
:SOURce2 <NRf>
:OFFSet
:DATE <yyyy>,<mm>,<dd>
:NDUE <yyyy>,<mm>,<dd>
:SAVE
:LOCK
Calibration subsystem.
Calibration commands protected by password.
Unlock calibration. (Default password: KI002500.)
Calibrate active range of channel 1 current measurement.
Calibrate active range of channel 2 current measurement.
Calibrate active channel 1 voltage bias range.
Calibrate channel 1 input voltage offset.
Calibrate active channel 2 voltage bias range.
Calibrate channel 2 input voltage offset.
Program calibration year, month, day.
Program calibration due year, month, day.
Save calibration data in EEPROM.
Lock out calibration.
Summary of Contents for 2500
Page 2: ...Contains Calibrating and Servicing Information Model 2500 Dual Photodiode Meter Service Manual...
Page 11: ......
Page 13: ......
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 28: ...2 Calibration...
Page 56: ...3 RoutineMaintenance...
Page 59: ...3 4 Routine Maintenance Model 2500 Service Manual...
Page 60: ...4 Troubleshooting...
Page 74: ...5 Disassembly...
Page 81: ...5 8 Disassembly Model 2500 Service Manual...
Page 82: ...6 ReplaceableParts...
Page 93: ...6 12 Replaceable Parts Model 2500 Service Manual...
Page 94: ...A Specifications...
Page 98: ...B CalibrationReference...
Page 108: ...C CalibrationProgram...
Page 117: ......
Page 119: ......
Page 120: ......
Page 121: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...