2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
KIPulse
The KIPulse factory script provides examples of how to generate pulses and to provide a simple
pulsing interface. Pulses can be generated using the functions listed below.
NOTE This factory script only operates on the channels present in the
SourceMeter instrument executing the pulse functions. These
functions will not operate correctly if the user attempts to access
SourceMeter instrument channels over the TSP-Link.
NOTE
PulseIMeasureV
and
PulseVMeasureI
may be accessed from the
front panel. The remaining scripts may only be accessed remotely.
The “
Config
” functions are used to configure a pulse train and assign the configuration to the
tag
parameter. The
Initiate
functions are used to execute the pulse train(s) assigned to its
tag
argument(s). The conditions listed in the table below must be true for these functions to execute
successfully.
Table 19-9
Required true conditions for “Initiate” function execution
Config
InitiatePulseTest
InitiatePulseTestDual
Source auto-range (I and V) off.
Output on.
Output on.
Measure auto-range (I and V) off.
Enough free space in buffer.
Enough free space in buffer.
Measure NPLC < ton.
Buffer
appendmode
on when
pulse train is >1 point.
Buffer
appendmode
on when
pulse train is >1 point.
Measure autozero OFF or ONCE.
Safety interlock engaged when
using the 200V range.
Safety interlock engaged when
using the 200V range.
Different unique SMUs for each
tag
.
Same NPLC setting for each
tag
.
Same
toff
for each
tag.