20-6
Return to
2600AS-901-01 Rev. B / September 2008
Section 20: Calibration
Series 2600A System SourceMeter® Instruments Reference Manual
Table 20-3
Model 2611A/2612A calibration steps
Function
1
Calibration steps
2
Calibration points
3
Sense mode
4
Voltage Source
and Measure
200mV
200mV
2V
2V
20V
200V
±1e-10, ±180mV
±1e-10, ±180mV
±1e-10, ±1.8V
±1e-10, ±1.8V
±1e-10, ±18V
±1e-10, ±180V
smuX.SENSE_LOCAL
smuX.SENSE_REMOTE
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
Current Source
and Measure
100nA
1
μ
A
10
μ
A
100
μ
A
1mA
1mA
10mA
100mA
1A
1.5A
10A
±1e-10, ±90nA
±1e-10, ±0.9
μ
A
±1e-10, ±9
μ
A
±1e-10, ±90
μ
A
±1e-10, ±0.9mA
±1e-10, ±0.9mA
±1e-10, ±9mA
±1e-10, ±90mA
±1e-10, ±0.9A
±1e-10, ±1.35A
±1e-10, ±2.4A
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
1. Calibrate only the source for the SENSE_CALA sense steps.
2. Steps must be performed in the order shown.
3. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-10.
Calibration polarities must also be set as shown in the procedures.
4. Output must be off before changing to the CALA sense mode.