2-18
Return to
2600AS-901-01 Rev. B / September 2008
Section 2: DUT Test Connections
Series 2600A System SourceMeter® Instruments Reference Manual
Figure 2-27
Model 2601A/2602A-1 connections for test circuit shown in
Figure 2-28
Model 2636A connections for test circuit shown in
Using shielding and guarding together
shows connections for a test system that uses a noise shield, a safety shield, and
guarding. The guard shields are connected to the driven guard (G) of the SMU. The noise shield is
connected to SMU LO. The safety shield is connected to the chassis and to a safety earth ground.
CHANNEL A
LO
G
LO
HI G G G HI
S
S
Model 2611A/2612A
DUT
HI
LO
Safety
Earth
Ground
Chassis
Chassis
Screw