2600AS-901-01 Rev. B / September 2008
Return to
9-13
Series 2600A System SourceMeter® Instruments Reference Manual
Section 9: Sweep Operation
3.
Request readings.
Request readings from Buffer 1 as follows:
printbuffer(1, 10, smua.nvbuffer1.readings)
Pulse sweep example
1.
Configure source functions
Examples:
The following commands restore defaults and set the compliance to 10V:
-- Restore Series 2600A defaults.
smua.reset()
--Set compliance to 10V.
smua.source.limitv = 10
2.
Configure and execute the sweep.
Example:
The following parameters configure a 10mA current pulse sweep with a 10ms
pulse on time, a 50ms pulse off time, and 10 pulse-measure cycles:
-- Pulse current sweep, Channel A, 1mA bias, 10mA level, 10ms pulse on,
-- 50ms pulse off, 10 cycles.
PulseIMeasureV(smua, 1e-3, 10e-3, 20e-3, 50e-3, 10)
3.
Request readings.
Request readings from Buffer 1 as follows:
printbuffer(1, 10, smua.nvbuffer1.readings)
List sweep example
1.
Configure source functions
Examples:
The following commands restore defaults and set the compliance to 10mA:
-- Restore Series 2600A defaults.
smua.reset()
-- Set compliance to 10mA.
smua.source.limiti = 10e-3
2.
Configure and execute the sweep.
Example:
The following parameters configure a list sweep with 3V, 1V, 4V, 5V, 2V points
using a 0.1s settling time:
-- Define voltage list.
vlist = {3, 1, 4, 5, 2}
-- List sweep, channel A, 3V, 1V, 4V, 5V, 2V steps, 0.1s delay, 5
-- points.
SweepVListMeasureI(smua, vlist, 0.1, 5)
3.
Request readings.
Request readings from Buffer 1 as follows:
printbuffer(1, 5, smua.nvbuffer1.readings)