2600AS-901-01 Rev. B / September 2008
Return to
10-5
Series 2600A System SourceMeter® Instruments Reference Manual
Section 10: Triggering
Figure 10-2
Remote trigger model
Yes
No
Another
arm?
Another
trigger?
No
Yes
End sweep
action
Arm
event
detector
smuX.trigger.arm.stimulus
smuX.trigger.source.stimulus
smuX.trigger.measure.stimulus
smuX.trigger.endpulse.stimulus
Sweeping event
smuX.trigger.SWEEPING_EVENT_ID
Armed event
smuX.trigger.ARMED_EVENT_ID
Source complete event
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
Measure complete event
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
Pulse complete event
smuX.trigger.PULSE_COMPLETE_EVENT_ID
Sweep complete event
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
Idle event
smuX.trigger.IDLE_EVENT_ID
Source
event
detector
Source
action
Measure
event
detector
Measure
action
End pulse
event
detector
End pulse
action
End pulse event overrun
Measure event overrun
Source event overrun
Arm event overrun
Trigger layer
Arm layer
Idle