11-2
Return to
2600AS-901-01 Rev. B / September 2008
Section 11: Display Operations
Series 2600A System SourceMeter® Instruments Reference Manual
Display functions and attributes
The display functions and attributes are used to perform the display operations covered in this
section.
lists each display function/attribute (in alphabetical order) and cross references
it to the section topic where the function/attribute is explained.
provides additional information on the display functions and attributes.
Display features
Display screen
The Keithley Instruments Series 2600A System SourceMeter
®
instrument can display source-
measure values and readings or user defined messages. The display screen options include the
following:
•
Source-measure, compliance screens:
• Display source and compliance values, and measure readings for SMU A.
• Display source and compliance values, and measure readings for SMU B (Models 2602A/
2612A/2636A only).
• Source-measure screen – Display source values and measure readings for SMU A and SMU
B (Models 2602A/2612A/2636A only).
•
User screen – Display user-defined messages and prompts.
The
display.screen
attribute is used to select the display screen:
display.screen = displayId
where:
displayId
is set to one of the following values or names:
0
or
display.SMUA
Table 11-1
Cross referencing functions/attributes to section topics
Function/Attribute
Section Topic
display.clear
display.getannunciators
display.getcursor
display.gettext
display.input
display.inputvalue
display.loadmenu.add
display.loadmenu.delete
display.locallockout
display.menu
display.prompt
display.screen
display.sendkey
display.setcursor
display.settext
display.smuX.digits
display.smuX.measure.func
display.trigger.clear
display.trigger.wait