19-82
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
lan.trigger[N].ipaddress
Replace N with values 1-8.
Attribute
Sets the IP address for outgoing trigger events.
Default
“0.0.0.0”
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
ipaddress = lan.trigger[lanevent].ipaddress
lan.trigger[lanevent].ipaddress = ipaddress
ipaddress
Sets and stores the LAN address for this attribute.
lanevent
The LAN event number.
Remarks
•
ipaddress
must be a string in dotted decimal notation.
• After changing this setting,
lan.trigger[n].connect
must be called before outgoing
messages can be sent.
lan.trigger[N].mode
Replace N with values 1-8.
Attribute
Sets the trigger event operation/detection mode
.
Default
lan.TRIG_EITHER
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
mode = lan.trigger[lanevent].mode
lan.trigger[lanevent].mode = mode
lanevent
The LAN event number.
mode
The trigger mode.
Choose one the following values for
mode
:
lan.TRIG_EITHER
Input. Detects rising or falling edge trigger packets
Output. A negative state LAN trigger packet.
lan.TRIG_FALLING
Input. Detects falling edge trigger packets.
Output. A negative state LAN trigger packet.
lan.TRIG_RISING
Input. Detects rising edge trigger packets.
Output. A positive state LAN trigger packet.
lan.TRIG_RISINGA
Input. Detects rising edge trigger packets.
Output. A positive state LAN trigger packet. Same
as
lan.TRIG_RISING
.
lan.TRIG_RISINGM
Input. Detects rising edge trigger packets.
Output. A positive state LAN trigger packet. Same
as
lan.TRIG_RISING
.
lan.TRIG_SYNCHRONOUS
Input. Detects falling edge trigger packets.
Output. A positive state LAN trigger packet.
lan.TRIG_SYNCHRONOUSA
Input. Detects falling edge trigger packets.
Output. A positive state LAN trigger packet.
lan.TRIG_SYNCHRONOUSM
Input. Detects rising edge trigger packets.
Output. A negative state LAN trigger packet.
Remarks
• lan.TRIG_EITHER
is the default mode.
• The
lan.TRIG_SYNCHRONOUS
values are provided for compatibility with digio and tsplink
triggering.