19-118
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
smuX.measure.interval
X = SMU channel (a or b)
Attribute
Interval between multiple measurements.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
interval = smuX.measure.interval
-- Reads measure interval.
smuX.measure.interval = interval
-- Writes measure interval.
interval
Set interval (in seconds) from 0 to 1.
Remarks
• This attribute sets the time interval between groups of measurements when
smua.measure.count
is set to a value greater than 1. The SMU will do its best to start the
measurement of each group when scheduled.
• If filtered measurements are being made, this interval is from the start of the first measurement
for the filtered reading to the first measurement for a subsequent filtered reading. Extra
measurements made to satisfy a filtered reading are not paced by this interval.
• If the SMU cannot keep up with the interval setting, measurements will be made as fast as
possible.
• The
reset
function sets the measure interval to 0.
Details
See
.
Also see
,
Example
Sets measure interval for SMU A:
smua.measure.interval = 0.5
smuX.measure.lowrangeY
X = SMU channel (a or b)
Y = SMU measure function (
v
or
i
)
Where:
v
= voltage,
i
= current
Attribute
Lowest measure range that will be used during autoranging.
Default
100e-9 (2601A/2602A/2611A/2612A)
100e-12 (2635A/2636A)
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
rangeval = smuX.measure.lowrangev -- Reads voltage low range.
smuX.measure.lowrangev = rangeval -- Writes voltage low range.
rangeval = smuX.measure.lowrangei -- Reads current low range.
smuX.measure.lowrangei = rangeval -- Writes current low range.
rangeval
Set to the lowest voltage or current measure range.
Remarks
• This attribute is used with auto-ranging to put a lower bound on the range used. Lower ranges
generally require greater settling times. By setting a low range value, measurements might be
able to be made with less settling time.
• If the instrument is set to auto range and it is on a range lower than the one specified, the range
will be changed to the range specified.
Details
See
Also see
Example
Sets volts low range for Model 2601A/2602A SMU A to 1V:
smua.measure
.
lowrangev
=
1