19-162
Return to
2600AS-901-01 Rev. B / September 2008
Section 19: Remote Commands
Series 2600A System SourceMeter® Instruments Reference Manual
status.operation.condition
status.operation.enable
status.operation.event
status.operation.ntr
status.operation.ptr
Attribute
Operation status register set.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
Reads condition, enable, event, NTR, and PTR registers:
operreg = status.operation.condition
operreg = status.operation.enable
operreg = status.operation.event
operreg = status.operation.ntr
operreg = status.operation.ptr
Writes to enable, NTR, and PTR registers:
status.operation.enable = operreg
status.operation.ntr = operreg
status.operation.ptr = operreg
Set
operreg
to one of the following values:
0
status.operation.CALIBRATING
status.operation.CAL
status.operation.SWEEPING
status.operation.SWE
status.operation.MEASURING
status.operation.MEAS
status.operation.TRIGGER_OVERRUN
status.operation.TRGOVR
status.operation.REMOTE_SUMMARY
status.operation.REM
status.operation.USER
status.operation.INSTRUMENT_SUMMARY
status.operation.INST
status.operation.PROGRAM_RUNNING
status.operation.PROG
Clears all bits.
Sets CAL bit (B0).
Sets CAL bit (B0).
Sets SWE bit (B3).
Sets SWE bit (B3).
Sets MEAS bit (B4).
Sets MEAS bit (B4).
Sets TRGOVR bit (B10).
Sets TRGOVR bit (B10).
Sets REM bit (B11).
Sets REM bit (B11).
Sets USER bit (B12).
Sets INST bit (B13).
Sets INST bit (B13).
Sets PROG bit (B14).
Sets PROG bit (B14).
operreg
can also be set to the decimal weight of the bit to be set. Examples:
To set bit B0 (CAL), set
operreg
to 1 (2
0
).
To set bit B4 (MEAS), set
operreg
to 16 (2
4
).
To set bit B11 (REM), set
operreg
to 2048 (2
11
).
To set more than one bit of the register, set
operreg
to the sum of their decimal weights.
For example, to set bits B0 and B4, set
operreg
to 17 (1 + 16).