tsplink
tsplink.group
Attribute
Stores the group number assigned to the node.
Default
0
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
groupnumber = tsplink.group
tsplink.group = groupnumber
groupnumber
The TSP-Link group number.
Remarks
• The default group number is 0.
• Set the attribute value to 0 to remove the node from all groups.
• Each time the node powers off, the group number for that node changes to 0.
• Group numbers can range from 0 to 64.
tsplink.master
Attribute
Reads the node number assigned to the master node.
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
master = tsplink.master
master
A custom variable that stores the master node number.
Remarks
This is a read-only attribute.
tsplink.node
Attribute
Use this attribute to set the node number.
Default
1
TSP-Link
accessibility
This attribute can be accessed from a remote TSP-Link node.
Usage
mynode = tsplink.node
Reads the node number.
tsplink.node = mynode
Writes the node number.
mynode
Sets the node number. Choose from 1 to 64.
Remarks
• This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
• Changes to the node number do not take effect until the next time
tsplink.reset
is executed
on any node in the system.
• Each node connected to the TSP-Link must be assigned a different node number.
• 32 nodes is the maximum number of nodes allowed.
• This attribute may be any number between 1 and 64.
Details
See
Also see
,
Example
Sets the TSP-Link node to number 2:
tsplink.node = 2