Section 4: Source-Measure Concepts
Series 2600A System SourceMeter® Instruments Reference Manual
Overview
The documentation in this section provides detailed information on source-measure concepts and
includes the following information:
•
•
•
•
•
Compliance limit
When sourcing voltage, the Keithley Instruments Series 2600A System SourceMeter
®
instrument
can be set to limit current. Conversely, when sourcing current, the SourceMeter instrument can be
set to limit voltage. The Series 2600A output does not exceed the compliance limit, except for the
compliance limit conditions described in
.
Maximum compliance
The maximum compliance values for the source ranges are summarized in
.
Compliance principles
Compliance acts as a clamp. If the output reaches the compliance value, the SourceMeter
instrument attempts to prevent the output from exceeding that value. This action implies that the
Table 4-1
Maximum compliance limits
Model 2601A/2602A
Model 2611A/2612A
Model 2635A/2636A
Source
range
Maximum
compliance
value
Source
range
Maximum
compliance
value
Source
range
Maximum
compliance
value
100mV
1V
6V
40V
3A
3A
3A
1A
200mV
2V
20V
200V
1.5A
1.5A
1.5A
100mA
200mV
2V
20V
200V
1.5A
1.5A
1.5A
100mA
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
3A
40V
40V
40V
40V
40V
40V
40V
40V
6V
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
1.5A
200V
200V
200V
200V
200V
200V
200V
20V
20V
100pA
1nA
10nA
100nA
1µA
10µA
100µA
1mA
10mA
100mA
1A
1.5A
200V
200V
200V
200V
200V
200V
200V
200V
200V
200V
20V
20V