Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-112
2651A-901-01 Rev. A / March 2011
lan.linktimeout
This attribute contains the LAN link timeout period.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
LAN restore defaults
Nonvolatile memory 20
Usage
timeout
= lan.linktimeout
lan.linktimeout =
timeout
timeout
The LAN link monitor time-out period (in seconds).
Details
You must enable the command
lan.autoconnect
before you can use this attribute.
The
timeout
value represents the amount of time that passes before the instrument disconnects from the LAN
due to the loss of the LAN link integrity.
The LAN interface does not disconnect if the connection to the LAN is reestablished before the
timeout
value
expires.
If the LAN link integrity is not restored before the
timeout
value expires, the instrument begins to monitor for a
new connection.
Example
print(lan.linktimeout)
Outputs the present LAN link timeout setting.
Also see
(on page 7-104)