Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-191
smuX.measure.autorangeY
This attribute stores the measurement autorange setting.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall
Saved setups
smuX.AUTORANGE_ON
Usage
autoRange
= smu
X
.measure.autorange
Y
smu
X
.measure.autorange
Y
=
autoRange
autoRange
The state of the measurement autorange setting; set to one of the following values:
0
or
smu
X
.AUTORANGE_OFF
: Disabled
1
or
smu
X
.AUTORANGE_ON
: Enabled
2
or
smu
X
.AUTORANGE_FOLLOW_LIMIT
: Measure range automatically set to the
limit range.
X
Source-measure unit (SMU) channel (for example,
smua.measure.autorangev
applies to SMU channel A).
Y
SMU measure function (
v
= voltage,
i
= current)
Details
This attribute indicates the measurement autorange state. Its value will be
smuX.AUTORANGE_OFF
when the
SMU measure circuit is on a fixed range and
smuX.AUTORANGE_ON
when it is in autorange mode.
Setting this attribute to
smu
X
.AUTORANGE_OFF
puts the SMU on a fixed range. The fixed range will be the
present SMU measure range.
Setting this attribute to
smu
X
.AUTORANGE_ON
puts the SMU measure circuit into autorange mode. It will remain
on its present measure range until the next measurement is requested.
If source highC mode is enabled, current autorange will be set to
smu
X
.AUTORANGE_FOLLOW_LIMIT
and
cannot be changed.
Example
smua.measure.autorangev = 1
Enables voltage measurement autoranging for SMU
channel A. Alternatively, the value 1 may be replaced with
smua.AUTORANGE_ON
.
Also see
(on page 7-209)