Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-196
2651A-901-01 Rev. A / March 2011
smuX.measure.delayfactor
This attribute stores a multiplier to the delays used when
smu
X
.measure.delay
is set to
smuX.DELAY_AUTO
.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Not saved
1
Usage
delayFactor
= smu
X
.measure.delayfactor
smu
X
.measure.delayfactor =
delayFactor
delayFactor
The delay factor multiplier.
X
Source-measure unit (SMU) channel (for example,
smua.measure.delayfactor
applies to SMU channel A)
Details
The delay factor is only applied when
smu
X
.measure.delay = smu
X
.DELAY_AUTO
.
This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic delay.
This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the automatic delay.
Setting this attribute to zero disables delays when
smu
X
.measure.delay = smu
X
.DELAY_AUTO
.
Example
smua.measure.delayfactor = 2.0
Doubles the measure delay for SMU channel A.
Also see
(on page 2-66)
(on page 7-209)
smuX.measure.filter.count
This attribute sets the number of measured readings required to yield one filtered measurement.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Recall
Saved setups
1
Usage
filterCount
= smu
X
.measure.filter.count
smu
X
.measure.filter.count =
filterCount
filterCount
The number of readings required for each filtered measurement. Set to a value
between 1 and 100.
X
SMU channel (for example,
smua.measure.filter.count
applies to SMU
Channel A).