Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-205
smuX.measure.rel.levelY
This attribute sets the offset value for relative measurements.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Instrument reset
SMU reset
Not saved
0
Usage
relValue
= smu
X
.measure.rel.level
Y
smu
X
.measure.rel.level
Y
=
relValue
relValue
Relative measurement offset value.
X
SMU channel (for example,
smua.measure.rel.levelv
applies to SMU Channel
A).
Y
SMU measurement function (
v
= voltage,
i
= current,
r
= resistance,
p
= power).
Details
This attribute specifies the offset value used for relative measurements. When relative measurements are
enabled (see
smu
X
.measure.rel.enable
Y
), all subsequent measured readings will be offset by the value of
this attribute. Specifically, each returned measured relative reading will be the result of the following calculation:
Relative reading = Actual measured reading - Relative offset value
Example
smua.measure.rel.levelv = smua.measure.v()
Performs a voltage measurement using
SMU Channel A and then uses it as the
relative offset value.
Also see
(on page 7-209)
smuX.measure.Y()