Section 7: Command reference
Model 2651A High Power System SourceMeter® Instrument Reference Manual
7-208
2651A-901-01 Rev. A / March 2011
Also see
(on page 7-242)
(on page 7-243)
(on page 7-245)
(on page 3-19)
smuX.nvbufferY
This attribute contains the dedicated reading buffer.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
See
Details
Not
applicable
Usage
bufferVar
= smu
X
.nvbuffer
Y
bufferVar
The dedicated reading buffer
X
Source-measure unit (SMU) channel (for example,
smua.nvbuffer1
applies to
SMU channel A)
Y
SMU nonvolatile buffer (
1
or
2
)
Details
Each SMU channel contains two dedicated reading buffers:
smu
X
.nvbuffer1
and
smu
X
.nvbuffer2
.
All routines that return measurements can also store them in either reading buffer. Overlapped measurements
are always stored in a reading buffer. Synchronous measurements return either a single-point measurement or
can be stored in a reading buffer if passed to the measurement command.
The dedicated reading buffers can be saved to internal nonvolatile memory (to retain data between power cycles)
using the
smu
X
.savebuffer()
function.
Example
smua.measure.overlappedv(smua.nvbuffer1)
Store voltage readings from SMU channel A
into SMU channel A dedicated reading
buffer 1.
Also see
(on page 7-201)
(on page 7-209)
(on page 7-235)
(on page 7-236)
(on page 7-238)
Configuring and running sweeps
(on page 3-29)
(on page 3-6)