Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-215
Details
Reading this attribute gives the output function of the source for the specified SMU Channel. Setting this attribute
configures the specified SMU Channel as either a voltage source or a current source.
Example
smua.source.func = smua.OUTPUT_DCAMPS
Selects the source current function
for SMU Channel A.
Also see
smuX.source.highc
This attribute sets high capacitance mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall
Saved setups
smuX.DISABLE
Usage
highC
= smu
X
.source.highc
smu
X
.source.highc =
highC
highC
The state of the high capacitance mode. Set to one of the following values:
0
or
smu
X
.DISABLE
: Disables high capacitance mode.
1
or
smu
X
.ENABLE
: Enables high capacitance mode.
X
SMU channel (for example,
smua.source.highc
applies to SMU Channel A).
Details
Turning on High-C mode has the following effects on the SMU settings:
smu
X
.measure.autorangei
is set to
smu
X
.AUTORANGE_FOLLOW_LIMIT
and cannot be changed.
Current ranges below 1
μ
A are not accessible
If
smu
X
.source.limiti
is less than 1
μ
A, it is raised to 1
μ
A.
If
smu
X
.source.rangei
is less than 1
μ
A, it is raised to 1
μ
A.
If
smu
X
.source.lowrangei
is less than 1
μ
A, it is raised to 1
μ
A.
If
smu
X
.measure.lowrangei
is less than 1
μ
A, it is raised to 1
μ
A.
Example
smua.source.highc = smua.ENABLE
Activates high capacitance mode
for SMU Channel A.
Also see