Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-219
Usage
offfunc
= smu
X
.source.offfunc
smu
X
.source.offfunc =
offfunc
offfunc
Set to the source function to be used when the output is off and the SMU is in
normal output-off mode. Set to one of the following values:
smu
X
.OUTPUT_DCVOLTS
: Source 0 V
smu
X
.OUTPUT_DCAMPS
: Source 0 A
X
SMU channel (for example,
smua.source.offfunc
applies to SMU channel A)
Details
This attribute controls the source function used when the output is turned off and the
smu
X
.source.offmode
is
set to
smu
X
.OUTPUT_NORMAL
. Set this attribute to
smu
X
.OUTPUT_DCVOLTS
for the source to be a 0 V source
when the output is off (
smu
X
.source.offlimiti
is used). Set it to
smu
X
.OUTPUT_DCAMPS
for the source to
be a 0 A source when the output is off (
smu
X
.source.offlimitv
is used).
This attribute is only used when the
smu
X
.source.offmode
attribute is set to
smu
X
.OUTPUT_NORMAL
.
Example
smua.source.offfunc = smua.OUTPUT_DCVOLTS
Sets the normal output-off mode to
source 0 V when the output is
turned off for SMU channel A.
Also see
(see "Output-off states" on page 2-59)
(on page 7-219)
(on page 7-220)
smuX.source.offlimitY
This attribute sets the limit (current or voltage) used when the source-measure unit (SMU) is in normal output-off
mode.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall
Saved setups
Current: 1e-3 (1 mA)
Voltage: 40 (40 V)
Usage
sourceLimit
= smu
X
.source.offlimit
Y
smu
X
.source.offlimit
Y
=
sourceLimit
sourceLimit
Set to the limit to be used when the SMU is in normal output-off mode
X
SMU channel (for example,
smua.source.offlimiti
applies to SMU channel A)
Y
SMU source function (
v
= voltage,
i
= current).
Details
Setting the current limit to lower than 1 mA may interfere with operation of the contact check function. See
(on page 7-186) and
(on page 7-187) for details.