Model 2651A High Power System SourceMeter® Instrument Reference Manual
Section 7: Command reference
2651A-901-01 Rev. A / March 2011
7-231
Example
smua.trigger.endpulse.action =
smua.SOURCE_IDLE
smua.trigger.endpulse.stimulus =
trigger.timer[1].EVENT_ID
Configure end pulse action to achieve a
pulse and configures trigger timer 1 to
control the end of pulse.
Also see
(on page 3-32)
smuX.trigger.endpulse.set()
This function sets the end pulse event detector to the detected state..
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
smu
X
.trigger.endpulse.set()
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.set()
applies to SMU channel A).
Details
This function sets the end pulse event detector to the detected state.
The SMU will automatically clear all the event detectors when the
smu
X
.trigger.initiate()
function is
executed. This function should be called after the sweep is initiated. Make sure that if the event detectors are
configured to automatically clear (because the
smu
X
.trigger.autoclear
attribute is set to
smu
X
.ENABLE
)
that this command is issued after the SMU has entered the trigger layer.
Also see
(on page 7-229)
(on page 7-234)
(on page 3-32)
smuX.trigger.endpulse.stimulus
This attribute defines which event will cause the end pulse event detector to enter the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Not saved
0